+86-15241797028
export@txinstruments.com
technique articles

Calibration of MS-2a Brinell optical scan system

The calibration procedure for MS-2a Brinell Indentation Measurement System is very simple.

Click “calibration” button, the condition for both the diameter of the indentation and 

measurement value will change into gray color on the right of software. After calibration, 

click “finish” button. (As shown in figure 9)

微信图片_20220901151502.png


Changing the diameter of indentation or the hardness value both methods can realize calibrate 

function.

图片1.png


Notice: When calibrate the hardness value or indentation diameter, please use dot marks, 

DO NOT use commas. I shall use 2.5mm/187.5kgf test condition as a example below: 

First correct calibration:

first correct calibration.jpg

First wrong calibration with commas:

first wrong calibration.jpg

Second correct calibration:

second correct calibration.jpg

Second wrong calibration with commas:

second wrong calibration.jpg

Result: You can see there are big difference if calibrate the value with comma.